# How to calibrate the Folium CO₂ sensor

> To keep CO₂  readings accurate, calibrate the Folium CO₂ sensor outdoors every 6 months using these instructions below. 

<video controls className="w-full aspect-video rounded-xl" src="https://5939375.fs1.hubspotusercontent-na1.net/hubfs/5939375/Folium%20Setup%20-%208%20Calibrating%20the%20CO2.mp4" />

| **Step** | **Calibration**                                                                                                                    | **Notes**                                                                                                                                                                                                                                               |
| :------- | :--------------------------------------------------------------------------------------------------------------------------------- | :------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| 1        | Take the **Folium sensor**(s) outside and wait **5** minutes for the sensor(s) to stabilise.                                       | **Note:** The time is to allow for the sensor to start accurately reading it's new environment. **Avoid** breathing towards the unit for these steps as the additional CO₂ contained in exhaled human breath can cause the calibration to be incorrect. |
| 2        | Press and hold the Power Button on the bottom of the **Folium sensor** for approx. 30 seconds until the unit lights up **orange**. | **Note:** If the button is only pressed or is released early the unit will flash green and the user will have to wait until the flashing stops to restart the calibration.                                                                              |
| 3        | When the **orange** flashing finishes with a **green** flash the calibration is complete.                                          | **Note:** the calibration technique assumes atmospheric CO₂ is 400ppm and that the sensors CO₂ sensor is reading atmospheric conditions.                                                                                                                |
| 4        | Repeat calibration every 6 months.                                                                                                 | \*\* \*\*                                                                                                                                                                                                                                               |

<Danger>
  If you accidently release the power button during the calibration period, the lights will light up green. Please wait for this to stop and then try again.
</Danger>